System Requirements (as presented by Phil Zeman project advisor)
- 10 Channels of input (each capable of EEG or EMG) 
- 16 bit quantization (minimizing quantization distortion) 
- Sample rate: 8000 10000 samp/sec/chan 
- Input impedance: 10MOhms 
- CMRR > 80 dB 
- Bandwidth: 0.1 to 1000Hz 
- Input range: EEG(1 160 μVpp) EMG(1 100 mVpp ) 
- Gain: EEG gain = 86 110dB 
- EMG gain = 30 70dB 
- Signal resolution: EEG(90nV steps) EMG(92μV steps) 
 